Network Settings

  • No network - messages are routed directly from Server to Client. TCPIP channel is not used.
  • Server- Server and Client are started at the same computer. Client can be disconnected if needed.
  • Client - Only Client is started. This is the mode of a new installation (except the demo version, which starts in No network mode): switch the spectrometer computer to Server or No network.

Access to the Server (

Introduced in version 3.8.13

see Network Security):

  • TCPIP address - in Client mode, the address of the Server computer. The Client-Server port is fixed at 3128.
  • Console port - on: the Server opens the text console (Remote control) on port 8023 (fixed). Off (default): no console port. Takes effect after a restart.
  • Listen address - network address the Server accepts connections on. Empty: 127.0.0.1 when the list of allowed hosts names only this computer, all network cards otherwise. 0.0.0.0 means all network cards.
  • List of allowed hosts - computers allowed to connect: an IP (192.168.1.20), localhost, or a subnet (192.168.1.0/24). An empty list admits nobody.
  • List of allowed folders - folders that files may be loaded from and saved to, in addition to the data, template, configuration, pattern, calibration and data buffer folders.
  • Allow MATLAB - off: the Server runs no MATLAB code.
  • Allow Python - off: the Server runs no Python code.
  • Allow programs - off: the Server runs no processing programs and no driver helper programs.
  • Console encryption - 0 Off, 1 Allowed, 2 Required (default): whether the console port accepts plain and/or encrypted connections. The Client-Server port is always encrypted.
  • Connection password - password clients must give to connect; stored as a hash (the field shows the hash). Empty: no password.
  • Console login - console connections must send .s.login='password' first.

Max trace length and resolutions option describes the data compression of traces shown in the Scope and Viewport. It optimizes the data transfer between Server and Clients.

  • X Axis compression (checkbox to apply, how many points to leave)
  • Data compression (how many bits per point transfer)